Gas Chromatography (GC) Comparision

SpecificationModelThermo Scientific TRACE GC UltraThermo Scientific Trace 1300 seriesPerkinElmer Clarus 600Shimadzu GC-2010Shimadzu GC-2010 PLUS (HIGH-END GC)Shimadzu GC- 2014Buck Scientific Inc. 910 / 310Agilent Technologies 7820 GC
FeaturesUltra Fast GC : Only for SSL/FID or PTV-FID configurations. Heat up rate 1200 ?C/min linear throughout entire Temperature range.-Majorly useful for QC/QA and Routine Laboratories
-Productivity solutions for user's Needs ( Model 1310 GC useful for larger routine labs and Method Development ; while 1300 GC useful for basic routice labs and it is budgeted also. -Instant Connect Injectors-Detectors Module- take less than 2 min to change -Instant Connect Helium Saver Module - Patented S/SL injector reduce He gas consumption to using only capillary column and Nitrogen is used to all injections like Inlet purge, septum, split flow and vaporization.
Powered by the fastest available heat-up and cool-down conventional oven Shorten injection-to-injection time with a high-performance ovenIt contains Advanced Flow Technology which helps to explore multi range of application of GC by multidimentional GC (AFC, Advanced Flow Control) , Capillary backflush , Detector splitting system and AFT software. High- End module of GC-2010 Plus.Automated analyzes with use of PeakSimple Software. Optional Dual Column oven. Choice of 12 different detectors. 910 GC perform routine industrial analyses and complicated research separtations. While 310 GC ideal for operation in field E.P.A. and A.S.T.M.Optional ALS Supports one 7650A autoinjector with capacity for 50 sample vials or Supports one 7693A autoinjector with capacity for 16 sample vials
General Specifications-Dimensions (H × W × D): 45 × 44 × 67 cm
-Weight: 35 kg main unit plus 0.8 kg each Instant Connect injector or detector module
-Dimensions (H × W × D) Clarus 600 GC with Autosampler: 83 x 67 x 82 cm
-Weight Clarus 600 GC with Autosampler: 69 kg
-Dimensions (H × W × D): 51.5 x 44 x 53 cm
-Weight : 30 kg
-Dimensions (H × W × D):
-Weight :
-Dimensions (H × W × D): 40 × 69 × 60.7 cm
-Weight : 48kg
-Dimensions (H × W × D): 45.7 x 35.5 x 33 cm
-Weight : 18-32 kg
-Dimensions (H × W × D): 49 x 56 x 51 cm
-Weight : 50 kg
Column OvenSize (H x W x D) and L27 x27 x 17.7 cm ; 12.9 m22.9 x 22.9 x 24.9 cm28 x 28 x 17.5 cm ; 13.725 × 36 × 17.5 cm19.05 x 20.32 x 7.62 cm28.0 × 30.5 × 16.5 cm
Temp Rangeambient temp upto 4500Cambiebt +3 0C to 450 oC(Ambient + 4°C) to 400°C (using liquid CO2 gas: -50°C to 450°C)(Ambient + 4°C) to 400°C (using liquid CO2 gas: -50°C to 450°C)(Ambient + 10°C) ~ 400°C (using liquid CO2 gas: -50°C - 400°C)Ambient to 450 oC8 °C above ambient to 425 °C
Max. Temp Ramp1200C/min125 0C / min140 ?C/min2500C/min2500C/min250°C/min75 °C/min
Typical Cool Down4500 to 500 in 250 sec4500 to 500 in less than 4 min450 °C to 50 °C in less than 2 minutes450°C to 50°C in appro. 6 min max.450°C to 50°C in 3.4 min.300°C ~ 50°C in 6 min max. (at 25°C ambient temperature)4000C to 500C in less than 5 mins
Sub-ambient -90°C with liquid N2, -55°C with CO2 options -100°C with liquid N2, -50°C with CO2 options
InjectorsVaporizing InletsExtended Performance
Split-Splitless Injector ; Packed and Purged Packed Injector ; BEST PTV™ (Brightly Enhanced
Sample Transfer PTV Injector ;
Max Two Injectors installed: options Instast Connet S/SL Injector Instant Connet Progammable Temp. Vaporizer Injector Instant Connet He Saver S/SL Injector Instant Connect Gas Sampling ValvePacked-column injector; Split/splitless capillary injector; PSS – programmable split/splitless capillary injector; GAS-SAMPLING VALVESSplit/Splitless Injection, Direct InjectionSplit/Splitless Injector : SPL-2010 Plus
On-Column/Programmed Temperature Vaporization Injector : OCI/PTV-2010 Direct Injection Unit : WBI-2010 Plus
For Packed / Dual,For Capillary Split/Splitless, Direct
(Split/splitless injection mode),(Pressure mode direct injection),(Flow-rate mode direct injection),For Single Packed, Direct (AMC)
On-Column injector, heated S/SL, 10 port Gas sampling valve and heated valve, 42-vial liquid and 20-vial Internal Auto sampler, Heated static Headspace InjectorMaximum inlets installed two : Purged packed (PP)
Split/splitless capillary (S/SL)
Non-Vaporizing InletsTrue Cold On-Column InjectorProgrammable on-column capillary injector
DetectorsFID (Flame Ionization Detector)MDL : 2 pg C/s Linear Range : 106 Typical Applications -hydrocarbons
-almost all organic compounds, such as
alcohols, essential oils, and fatty acids
Available as Instant Connect,user-exchangeable modules. Maximum number installed: 3 including a mass spectrometers on the same oven, or 5 plus an MS when using the TRACE 1310 Auxiliary Oven
Minimum Detectable Level (MDL): <1.4 pg C/s
Linear range: >107 (±10%)
Maximum temperature: 450 °C in steps of 0.1 °C
Operating temperature :100 ?C to 450 ?C in 1 ?C increments
Minimum Detectable Level (MDL): < 3 -10-12 g C/sec nonane at a S/N = 2 to 1
Linearity : > 106
Makeup gas Not required

Temperature range : upto 450°C max.
Minimum detected quantity: 3pg C/s (dodecane)
Linearity range: 107




FID-2010 Plus

Minimum detected quantity: 1.5 pg C/s
Automatic ignition, re-ignition, and flame extinguishing
functions
Feedback function reduces gas supply pressure to zero
when the hydrogen flame is extinguished.





Temperature range : 400°C max.
Minimum detected quantity: 3pgC/s (dodecane)
Linearity range: 107




Operate Temperature : 4000C ; MDL: from ~10 ppm ; Linearity : ~10% for C1 to C10 compound; Carrier Gas : HeliumMaximum operating temperature : 425 °C
MDL : < 3 pg carbon/s as tridecane
Linearity range : > 107 range with N2 carrier and 0.29-mm id jet
TCD (Thermal Conductivity Detector)MDL : 600 pg ethane/ml He carrier Linear Range : 104 Typical Applications -permanents gases
-organic and inorganic compounds not
detectable by FID such as carbon disulphide,
ammonia, water
Maximum temperature: 400 °C in steps of 0.1 °C
MDL: <400 pg tridecane/mL with He carrier or <20 pg tridecane/s with a total flow through the cell of 3 mL/min Linearity range: 105
Operating temperature: 100 ?C to 350 ?C in 1 ?C increments
Minimum Detectable Level (MDL): Typically < 1 ppm nonane
Linearity : > 105
Temperature range : 400°C max.
Linearity range : 105



TCD-2010 Plus

Sensitivity : 20000 mV mL/mg (decane)
For analysis of gases and concentrated organic
compounds

Temperature range : 400°C max.
Linearity range : 105



Operate Temperature : 2750C ; MDL: ~250 ppm ; Linearity : ~100 % ; Carrier Gas : HeliumMaximum operating temperature 400 °C
MDL : < 800 pg propane/mL using He carrier
(MDL may be affected by laboratory
environment)
Linearity range : 105 (± 10%)
ECD (Electron Capture Detector)MDL : 0.01pg/s (lindane) Linear Range : 104 Typical Applications -organochlorinated pesticides
-halogenated contaminants in water
and soil
-PCBs in oil
MDL: <4.5 fg/s lindane
Linearity range: >104 with lindane
Maximum temperature: 400 °C in 0.1 °C steps
Make-up gas: Nitrogen or 95% argon/5% methane
Carrier gas Either Ar/CH4 or N2
Operating temperature : 100 ?C to 450 ?C in 1 ?C increments
Minimum Detectable Level (MDL): < 0.05 pg perchloroethylene with argon/methane or nitrogen
Linearity : > 104
Makeup gas Standard
ECD- 2010 Plus MDL: 6 fg/sec A highly sensitive and selective detector for the analysis of electophilic compounds.Temperature range : 400°C max.
Minimum detected quantity : 0.1pg/s ( -BHC) Linearity range : 104



MDL: 1-5 ppb ;Makeup gas types argon/5% methane or nitrogen
Radioactive source < 15 mCi 63Ni
MDL : < 0.02 pg/mL lindane
Linearity range: > 104 with lindane
NPD (Nitrogen Phosphorus Detector)MDL : N 0.05 pg/s; P 0.02 pg/s Linear Range : 104 Typical Applications -triazine herbicides
-organophosphorus pesticides (EPA 507)
-nicotine, caffeine nitrosamines
(EPA 607)
-NPD available with ceramic beads and compatible with element-specific sources
MDL: <20 fg P/s and <100 fg N/s with standard ceramic bead
Linearity range: 104
Maximum temperature: 450 °C in steps of 0.1 °C
Operating temperature : 100 ?C to 450 ?C in 1 ?C increments
Minimum Detectable Level (MDL): 5 -10-13 g N/sec 2,4-dimethylaniline, 5 -10-14 g P/sec tributylphosphate
Linearity : > 104
Makeup gas Not required
MDL: ~100 ppb for Nitrogen containing compounds and ~500ppb for phosporus containing compounds; Carrier Gas : HeliumMaximum operating temperature 400 °C
MDL : < 0.4 pg N/s, < 0.2 pg P/s with
azobenzene/malathion mixture

Linearity range : > 104 N, > 104 P with
azobenzene/malathion mixture
PID (Photo-Ionization Detector)MDL : Benzene 1 pg/s (using 10.6 eV UV lamp) Linear Range : 105 Typical Applications -low boiling aromatics (BTEX)
-policyclic aromatic hydrocarbons
-amines
Operating temperature :100 ?C to 250 ?C in 1 ?C increments (can be set to 350 ?C for cleaning)
Minimum Detectable Level (MDL): : < 10 pg benzene
Linearity : > 107
Makeup gas Standard
MDL: ~10-50ppb ;
FPD (Flame Photometric Detector)MDL : P = 0.1 pgP/s; S = 5 pgS/s Linear Range : 104 (P) Typical Applications -organophosphorus pesticide
-phosphine
-sulphur gases
MDL: 100 fg P/s and 5 pg S/s (Methyl Parathion)
Linearity range: 104 (P), >103 (S)
Maximum temperature: 450 °C base temperature, 200 °C cell temperature; in steps of 0.1 °C
-Also available in dual wavelength version
Operating temperature 250 ?C to 450 ?C in 1 ?C increments
Minimum Detectable Level (MDL): : 1 -10-11 g S/sec thiophene, 1 -10-12 g P/sec tributylphosphate
Linearity : Sulfur 102 (log-log), Phosphorus 103
Signal filtration 50, 200, 800 msec
Makeup gas Not required
FPD-2010 Pus MDL: 55 fg/sec (Phosphorus compounds) ; 3 pgS/sec (Sulfur Compounds)Temperature range : 350°C max.
Linearity Range : P: 104
S: 103



MDL: ~1-10 ppb ; Carrier Gas : HeliumMaximum operating temperature : 250 °C MDL : < 200 fg P/s, < 6 pg S/s with
methylparathion
Linearity range : > 5 × 102 S, 104 P with
methylparathion

FTDFTD-2010 Plus MDL : 0.1 pgN/sec For analysis of organo nitrogen and phosphorus
compounds, such as residual pesticides.
Temperature range : 400°C max.
Linearity Range :N: 103
P: 103




Operate Temperature : 2750C ; MDL: ~250 ppm ; Linearity : ~100 % ; Carrier Gas : Helium
HID (Helium Ionization Detector)MDL : ~ 1-10 ppm ; Carrier Gas : Helium
DELCD (Dry Electro-Lytic Conductivity Detector)MDL : 10 ppb ; Carrrier Gas : helium
BIDBID-2010 Plus MDL : 1 pgC/sec The barrier discharge ionization detector (BID) is a highly sensitive device that creates ionization from a Helium-based, dielectric barrier discharge plasma. It can detect all organic and inorganic compounds, with the exception of He and Ne, with no difference in sensitivity.
PDD (Pulsed Discharge Detector)MDL : 0.2 pg/s methane Linear Range : 105 Typical Applications -permanents gases
-light hydrocarbons
-organic and inorganic compounds not detectable by FID such as carbon disulphide, ammonia, water
CCD (Catalytic Combustion Detector)Operate Temperature : < 2000C ; MDL : 500 ppm ; Carrrier Gas : helium
IEC ( Integrated Electronic Control) Gas Specificationsupto 18 channels
CertificationsInternational Electrotechnical Commission (IEC): 61010-1:2001 - 61010-2-010:2003 - 61010-2-081:2001 + A1:(2003)
CAN/CSA C22.2 No. 61010-1 and UL 61010-1
EuroNorm (EN): 61010-1:2001 - 61010-2-010:2004 - 61010-2-081:2002
Electromagnetic Compatibility (EMC) and Radio Frequency Interference (RFI): CISPR 11/EN 55011: Group 1 Class A and IEC/EN 61326-1:2006
- See more at: http://www.thermoscientific.com/en/product/trace-1310-gas-chromatograph.html#sthash.ND1jfhlb.dpuf
International Electrotechnical Commission (IEC): 61010-1:2001 - 61010-2-010:2003 - 61010-2-081:2001 + A1:(2003)
CAN/CSA C22.2 No. 61010-1 and UL 61010-1
EuroNorm (EN): 61010-1:2001 - 61010-2-010:2004 - 61010-2-081:2002
Electromagnetic Compatibility (EMC) and Radio Frequency Interference (RFI): CISPR 11/EN 55011: Group 1 Class A and IEC/EN 61326-1:2006
- See more at: http://www.thermoscientific.com/en/product/trace-1310-gas-chromatograph.html#sthash.ND1jfhlb.dpuf
Canadian Standards Association (CSA) C22.2 No. 61010
CSA/Nationally Recognized Test Laboratory (NRTL) UL61010
International Electrotechnical Commission (IEC) 61010
EuroNorm (EN) EN61010
Electromagnetic compatibility (EMC) and radio frequency
interference (RFI) regulation conformity
CISPR 11/EN 55011 Group 1, Class A
IEC/EN 61326
Designed and manufactured under a quality system registered
to ISO 9001. The Declaration of Conformity is available.